Cube L-1

The PRIMES Cube L-1 is ready to use in just moments. Ideal for validating your machine’s power output and identifying fluctuations, helping you maintain performance, perform adjustments, and ensure consistent part quality. A must-have for every machine operator and field service professional!


MEASUREMENT PARAMETERS
Power Range: 200 – 16,000 W
Wavelength Range: 900– 1,100 nm
Beam Diameter on Protective Window: 1 – 7 mm
Max. Power Density on Protective Window: 250 kW/cm²
Max Beam Diameter on Absorber: 45 mm
Irradiation Time: 0.1 – 2.0 s (depending on laser power)
Min. On/Off Times (Duty Cycle) for Pulsed Lasers: 50 μs (e.g. max. 10 kHz at 50% duty cycle)
Max. Laser Rise Time: < 1% of irradiation time
Energy Per Measurement (depending on beam diameter 2): 200-4000 J
Recommended Energy Per Measurement: 500 – 2,000 J
Total Duration Until Measurement Value Output: < 15 s
Nominal Measurement Frequency:
700 J: 1 cycle/min
4,000 J: 1 cycle/15 min

DEVICE PARAMETERS
Max. Absorber Temperature: 120 °C
Max. Angle of Incidence (Perpendicular to Inlet Aperture): ± 5°
Max. Beam Divergence (Full Angle) at Angle of Incidence up to 5°: 160 mrad
Max. Centered Tolerance: ± 2.0 mm
Accuracy: ± 3 %
Reproducibility: ± 1 %

SUPPLY DATA
Power Supply: Built-in lithium-ion battery, which can be changed via micro-USB port
Temperature range for charging the lithium-ion cell: 0 – 45°C

COMMUNICATION
Interfaces: Bluetooth, micro-USB
Software: Cube App and LaserDiagnosticsSoftware (LDS)

DIMENSIONS AND WEIGHT
Dimensions (L × W × H, without connectors): 92 × 97 × 110 mm
Weight (Approx.): 1700 g

1 The stated limit values are to be understood in correlation with the permitted maximum energy (E = P ∙ t).
2 Limiting the maximum energy as a function of the beam diameter serves to protect the device and prolongs its service life.