Raw Beam and Power Measurement

BeamMonitor BM+

Analyzing collimated CO₂ lasers or process-critical power density distributions of solid-state lasers.
Overview

The PRIMES BeamMonitor BM+ is a measuring device for performing beam diagnostics on unfocused, continuous irradiation of CO₂ and solid-state lasers with higher outputs. With the BeamMonitor BM+, the user can quickly and reliably measure beam parameters such as the spatial power density distribution. Analyzing and documenting the laser parameters makes it possible to optimize the processing procedure or identify an incorrect beam position or incorrect lens alignment.

50–25,000 W
Power Range
5–70 mm
Beam Diameter
100 mrad
Max. Beam Divergence
Specifications
BeamMonitor BM+ is offered in two models (BM+ 60 / BM+ 100S); values below are listed per model where they differ.
Measurement Parameters
ParameterSpecification
Power Range 50–25,000 W (lower/higher powers on request)
Wavelength Range 450–1,090 nm or 10,600 nm
Beam Diameter BM+ 60: 5–42 mm (450–1,090 nm), 10–42 mm (CO₂) | BM+ 100S: 10–70 mm
Min. Power Density (CO₂ Only) BM+ 60: 0.1 kW/cm² | BM+ 100S: 0.2 kW/cm²
Max. Power Density 10 kW/cm² (higher on request)
Max. Beam Divergence 100 mrad
Irradiation Time 2 s to continuous
A/D Conversion 16 bit
Nominal Measuring Frequency Linescan — BM+ 60: 25 Hz | BM+ 100S: 30 Hz
Determined Parameters
ParameterSpecification
Beam Position x, y Yes
Beam Dimensions x, y Yes
Power Density Distribution 2D, 3D
Linescan Yes
Device Parameters
ParameterSpecification
Working Range x-y BM+ 60: 60 × 60 mm | BM+ 100S: 100 × 100 mm
Measurement Window Sizes BM+ 60: 0.1 × 0.1 mm – 60 × 60 mm | BM+ 100S: 0.1 × 0.1 mm – 100 × 100 mm
Resolution 32 × 32 – 1,024 × 1,024 px
Rotation Speed of Measuring Tip 1,562 min⁻¹
Accuracy (Beam Diameter) ±5%
Reproducibility (Beam Diameter) ±3%
Supply Data
ParameterSpecification
Power Supply 24 V DC ±5%, max. 1.8 A
Communication
ParameterSpecification
Interfaces Ethernet, RS485
Dimensions and Weight
ParameterSpecification
Dimensions (L × W × H) BM+ 60: 316 × 212 × 83 mm | BM+ 100S: 436 × 292 × 83 mm
Weight (Approx.) BM+ 60: 9 kg | BM+ 100S: 10 kg
Features
Capabilities organized by functional domain.
Measurements

A rotating measuring tip samples the laser beam point by point while the mirror holder moves linearly through the beam to capture the entire beam profile. A 16-bit A/D converter and a resolution of up to 1,024 × 1,024 pixels allow exact analysis of even the smallest disturbances in the raw beam. The device is built for industrial production environments and can be used in any orientation.

  • Beam position
  • Beam dimensions
  • Beam symmetry
  • Power density distribution
Applications

Laser beams are “customized” for specific applications using telescopes and adaptive lenses, which change the beam diameter and divergence dynamically — for example to shift the focus position in laser cutting and welding systems. The BM+ is also suited to quality assurance and laser approval: deterioration of laser lenses is one of the main causes of problems in laser material processing. Diode and other solid-state lasers can be checked in the collimated range with relatively little effort.

Operation

The PRIMES LaserDiagnosticsSoftware (LDS) is the graphical interface for operating the BM+, with tools for analysis, display and preparation of measurement data, including a report function. Data is communicated over Ethernet. During measurement the entire beam leaves the BM+ again, so it should be absorbed by a PowerMonitor or another PRIMES beam absorber.

Models & Options

The BM+ is available with a 60 mm or 100 mm aperture (BM+ 60 and BM+ 100S), each for CO₂ or NIR lasers. As a rule, the aperture should be at least 1.4 times the laser beam diameter, so the beam edges do not heat the casing or limit the measuring range.

Request & Resources
View full Literature Library →
Website Contact Form