The PRIMES Cube L1 features an advanced optical front end and a wide measurement range from a few hundred watts up to 16 kW, making it suitable for demanding laser service tasks. Its calorimetric measurement system provides precise, reproducible results referenced to the actual measured value, while the specialized optical front end supports power densities up to 250 kW/cm² for accurate measurements in tight, high-power environments.
Measurement Parameters
| Parameter | Specification |
|---|---|
| Power Range | 200–16,000 W |
| Wavelength Range | 900–1,100 nm |
| Beam Diameter on Protective Window | 1–7 mm |
| Max. Power Density on Protective Window | 250 kW/cm² |
| Max Beam Diameter on Absorber | 45 mm |
| Irradiation Time | 0.1–2.0 s |
| Min. On/Off Times (Duty Cycle) for Pulsed Lasers | 50 µs |
| Max. Laser Rise Time | <1% of irradiation time |
| Energy Per Measurement | 200–4,000 J |
| Recommended Energy Per Measurement | 500–2,000 J |
| Total Duration Until Measurement Value Output | <15 s |
| Nominal Measurement Frequency | 700 J: 1 cycle/min; 4,000 J: 1 cycle/15 min |
Device Parameters
| Parameter | Specification |
|---|---|
| Max. Absorber Temperature | 120°C |
| Max. Angle of Incidence | ±5° |
| Max. Beam Divergence (Full Angle) | 160 mrad |
| Max. Centered Tolerance | ±2.0 mm |
| Accuracy | ±3% |
| Reproducibility | ±1% |
Supply Data
| Parameter | Specification |
|---|---|
| Power Supply | Built-in lithium-ion battery, changeable via micro-USB port |
| Temperature Range for Charging | 0–45°C |
Communication
| Parameter | Specification |
|---|---|
| Interfaces | Bluetooth, micro-USB |
| Software | Cube App and LaserDiagnosticsSoftware (LDS) |
Dimensions and Weight
| Parameter | Specification |
|---|---|
| Dimensions (L × W × H) | 92 × 97 × 110 mm |
| Weight (Approx.) | 1700 g |