Focus Measurement

FocusMonitor FM+

The established standard in high power laser beam diagnostics.
Overview

The PRIMES FocusMonitor FM+ analyzes and measures laser beam sources in material processing. The FM+ determines the beam properties of focused laser beams — their geometric dimensions, focal position, beam parameter product and beam quality factor — and its integrated z-axis automatically measures entire caustics over up to four Rayleigh lengths, making it easy to verify compliance with quality standards. An update to the original FocusMonitor, the FM+ has new electronics for current and future signal processing requirements, a 16-bit A/D converter and an Ethernet interface for fast, secure data exchange with computers or system controllers. Its mechanical design also allows upside-down installation without additional components.

30–25,000 W
Power Range
100–3,000 µm
Beam Diameter
120 mm
Working Range z
Specifications
FocusMonitor FM+ is offered as the standard FM+ and the FM+ HPD for very high power densities; values below are listed per model where they differ.
Measurement Parameters
ParameterSpecification
Power Range 30–25,000 W (higher power ranges on request)
Wavelength Range FM+: 400–12,000 nm | FM+ HPD: 515–545 nm and 1,000–1,100 nm
Beam Diameter FM+: 100–3,000 µm (up to 5,000 µm optionally) | FM+ HPD: 100–1,200 µm
Max. Power Density FM+: CO₂ (10.6 µm) 30 MW/cm², Nd:YAG (1,000–1,100 nm) 10 MW/cm², VIS (515–550 nm) 5 MW/cm², diode (800–1,000 nm) 1 MW/cm² | FM+ HPD: 50 MW/cm²
Max. Beam Divergence (Depending on Tip) FM+: NIR high div 200 mrad, CO₂ high power 240 mrad, diode 500 mrad | FM+ HPD: 120 mrad
Determined Parameters
ParameterSpecification
Focus Position x, y, z Yes
Focus Radius x, y Yes
Beam Quality Factor M² Yes
Divergence Angle Yes
Power Density Distribution 2D, 3D
Device Parameters
ParameterSpecification
Measurement Window x-y FM+: 8 × 8 mm (12 × 12 mm optionally) | FM+ HPD: 8 × 8 mm
Working Range z 120 mm (larger on request)
Resolution 32 × 32 – 1,024 × 1,024 px
Rotation Speed FM+: 1,875 / 3,750 / 7,500 min⁻¹ | FM+ HPD: 1,875 min⁻¹
Linescan Yes
Supply Data
ParameterSpecification
Power Supply 24 V DC ±5%, max. 3.5 A
Inert Gas (Water and Oil Free) FM+: helium, nitrogen or argon | FM+ HPD: helium, nitrogen, argon or compressed air
Inert Gas Pressure Typ. 0.5 bar
Communication
ParameterSpecification
Interfaces Ethernet, RS485
Dimensions and Weight
ParameterSpecification
Dimensions (L × W × H) FM+: 280 × 242 × 218 mm | FM+ HPD: 318 × 242 × 218 mm
Weight (Approx.) 8.5 kg
Features
Capabilities organized by functional domain.
Measurements
  • Focal dimensions
  • Position of the focal point in space
  • Power density distribution
  • Symmetry of the power density distribution
  • Diffraction index M² or K
  • Beam parameter product
How It Works

The FM+ is an opto-mechanically scanning measuring system. A small hole in its measuring tip (typically about 20 µm in diameter) lets through a small section of the laser beam, which two mirrors guide to a detector chosen for the laser power and wavelength. The high orbital velocity of the rotating tip allows analysis of high power densities, and the wide dynamic range of the A/D converter shows very low intensities with the same precision as the highest peaks — a requirement for automatic caustic measurement over at least four Rayleigh lengths in accordance with ISO 11146. The FM+ HPD variant uses the Fused Silica Sensor System (FS³) to measure power densities of up to 50 MW/cm².

Operation

The LaserDiagnosticsSoftware (LDS) enables fully automatic, semi-automatic or manual measurement of the beam distribution, beam position and beam dimensions over a fast Ethernet connection, with higher-resolution measurements and a freely selectable region of interest. The number of measurement planes is freely selectable (typically 11 to 21), and parameters such as beam waist diameter, divergence angle, Rayleigh length and M² are calculated from the results.

Measuring Tips & Detectors

The measuring tip and detector can be exchanged in a few minutes, with versions for different divergences and wavelengths. CO₂ lasers mainly use pyroelectric detectors such as the DFCM+ with CO₂ high power tips; the DFY-PS+ covers 400–1,100 nm and the DFIG-PS+ 1–2.1 µm. Because hardly any beam power is absorbed during measurement, the FM+ is ideally combined with a power meter such as the PRIMES PowerMonitor.

Request & Resources
View full Literature Library →
Website Contact Form