Overview
The HighPower-MSM-HighBrilliance determines beam parameters of focused laser beams of multi-kilowatt lasers in the range of 20 micrometres up to a millimetre directly in the process zone – even at full power. For this purpose, 95% of the laser power is transmitted by a beam splitter through the measurement optic, and absorbed. The remaining five per cent are further attenuated in the measurement optic and destroyed by internal water-cooled absorbers. A part beam with a few milliwatts of power is imaged on the CCD sensor in an enlargement.
up to 20 kW (multi mode)
Max. Laser Power
20–1,000 µm
Beam Diameter
1,025–1,080 nm
Wavelength Range
Specifications
MicroSpotMonitor MSM+ HighBrilliance is offered in two models (HB10 / HB20); values below are listed per model where they differ.
Measurement Parameters
| Parameter | Specification |
|---|---|
| Models | MSM+ HB10 / MSM+ HB20 |
| Max. Laser Power | HB10: 5 kW single mode, 10 kW multi mode | HB20: 10 kW single mode, 20 kW multi mode |
| Admissible Wavelength Range | 1,025–1,080 nm |
| Beam Diameter | 20–1,000 µm |
| Admissible Measuring Range | ±3 zR |
| Max. Input NA | 0.11 |
| Max. Working Range z | HB10: 120 mm | HB20: 40 mm |
Supply Data
| Parameter | Specification |
|---|---|
| Power Supply | 24 V DC ±5%, max. 1.8 A (0.4 A in standby) |
| Recommended Cooling Water Flow | 0.7 l/min per kW |
| Min. Cooling Water Flow | HB10: 5 l/min | HB20: 10 l/min |
| Cooling Water Temperature Tin | Dew point < Tin < 30 °C |
| Max. Water Inlet Pressure | 4 bar |
| Compressed Air Pressure | 0.5–1.0 bar (max. 2 bar) |
| Compressed Air Specification | ISO 8573-1:2010 [6:4:4] |
Communication
| Parameter | Specification |
|---|---|
| Interfaces | Ethernet, PRIMES-Bus RS485, safety circuit |
Dimensions and Weight
| Parameter | Specification |
|---|---|
| Dimensions (L × W × H, excl. Cables and Plugs) | HB10: 600 × 401 × 388 mm | HB20: 727 × 400 × 385 mm |
| Focal Position Above Base Plate | HB10: 320–400 mm | HB20: 390–410 mm |
| Weight (Approx.) | HB10: 35 kg | HB20: 42 kg |