Raw Beam and Power Measurement

LaserQualityMonitor LQM+

Measures the quality of cw, pulsed and USP laser beams with a simple push of a button.
Overview

The PRIMES LaserQualityMonitor LQM+ quickly and automatically characterizes the beam parameters of a beam source. The measuring system is compact, easy to align and can be expanded to meet individual requirements. With the LQM+, PRIMES offers an integrated solution for the quick and straightforward analysis of the beam properties of a beam source: the beam propagation of lasers from the UV to the NIR range is characterized and certified, and errors in the optical structure of resonators or beam imaging systems can easily be analyzed.

up to 500 W
Max. Laser Power
1–60
Beam Quality M²
100 fs–cw
Pulse Duration
Specifications
LaserQualityMonitor LQM+ is offered in three models (LQM+ 20 / LQM+ 200 / LQM+ 500); values below are listed per model where they differ.
Measurement Parameters
ParameterSpecification
Max. Laser Power (at 1,064 nm) LQM+ 20: 20 W | LQM+ 200: 200 W | LQM+ 500: 500 W (depends on wavelength, beam quality, raw beam diameter and pulse characteristics)
Pulse Duration 100 fs–cw
Wavelength Range 340–360 nm (UV), 515–545 nm (green), 1,030–1,090 nm (NIR)
Beam Dimensions Single mode: 1.5–9 mm | Multi mode: 1.5–15 mm
Beam Quality Factor M² 1–60
Max. Beam Divergence 10 mrad
Determined Parameters
ParameterSpecification
Power Density Distribution 2D, 3D
Supply Data
ParameterSpecification
Power Supply 24 V DC ±5%, max. 1.8 A
Cooling LQM+ 20 / LQM+ 200: air cooling | LQM+ 500: water cooling
Cooling Water Pressure LQM+ 500: 2 bar primary pressure with an unpressurized outflow, max. 4 bar
Min. Cooling Water Flow Rate LQM+ 500: 1.5 l/min
Cooling Water Temperature Tin Dew point < Tin < 30 °C
Communication
ParameterSpecification
Interfaces Ethernet
Dimensions and Weight
ParameterSpecification
Dimensions (L × W × H) LQM+ 20: 285 × 190 × 180 mm | LQM+ 200/500: 350 × 230 × 190 mm
Weight (Approx.) LQM+ 20: 10 kg | LQM+ 200/500: 18 kg
Features
Capabilities organized by functional domain.
Measurements
  • Beam waist diameter / beam dimensions
  • Waist position / focal position relative to the beam entry into the LQM+
  • Beam radius
  • Far field divergence
  • Rayleigh length
  • Divergence
  • Beam propagation factor K
  • Diffraction index M²
How It Works

Characterizing a collimated beam directly would need a measurement range of several meters, so ISO 11146 recommends measuring a caustic generated by focusing instead. The LQM+ creates this caustic inside the device and images the focused beam, enlarged, onto a CCD sensor through integrated attenuators and a lens. The LaserDiagnosticsSoftware determines the beam radius, position and orientation from each measured power density distribution, and repeating the measurement at various positions determines every parameter of the caustic — shortening the measurement from several meters to a few millimeters. The results are calculated back to the collimated beam in accordance with ISO 11146.

Operation

Two alternatives are available for operating the LQM+. Data is transferred to the PC via Ethernet.

  • PC-based LaserDiagnosticsSoftware (LDS) for manual and semi-automatic measurement of beam distribution, position and dimensions
  • Scripts for semi-automatic control of repetitive measurements in servicing, quality assurance and acceptance inspection
Models & Options
  • Power measurement on the absorber
  • Crimp holders for collimators with 35, 40 and 50 mm outer diameter
  • Collimator with 67 mm focal length up to 6 kW, for LLK-D and QBH fiber connectors
  • Measurement optics 1:1 and 5:1
  • Wavelength ranges 1,030–1,090, 515–545 and 340–360 nm
  • Neutral density glass filters OD1 to OD5