Focus Measurement

MicroSpotMonitor MSM+

The most versatile camera-based beam profiler: micro to macro, femto to cw, UV to NIR.
Overview

The PRIMES MicroSpotMonitor MSM+ is the ideal tool for checking, monitoring and approving very finely focused laser beams, such as those used in micro-material processing. It automatically measures and analyzes the spatial beam distribution around the focus at various z positions. The focused laser beam is the actual tool in most laser material processing, and regular measurements identify and log any changes in beam profile, focal position or beam dimensions — supporting both failure analysis and process optimization. The “plus generation” adds higher measuring rates, enhanced algorithms and new measuring modes, including fully automatic caustic measurement.

1 mW–200 W
Power Range
10–1,000 µm
Beam Diameter
up to 15×
Magnification
Specifications
Datasheet specifications for the PRIMES MicroSpotMonitor MSM+.
Measurement Parameters
ParameterSpecification
Power Range 1 mW–200 W (optional 500 W with water cooling)
Wavelength Range 340–360 nm, 515–545 nm and 1,030–1,090 nm
Beam Diameter 10–1,000 µm
Measurement Results
ParameterSpecification
Focus Position x, y, z Yes
Focus Radius x, y Yes
Beam Quality Factor M² Yes
Raw Beam Diameter at Focusing Element Yes
Beam Parameter Product BPP Yes
Divergence Angle Yes
Power Density Distribution 2D, 3D
Device Parameters
ParameterSpecification
Travel Range x-y Plane 2 × 2 mm
Travel Range z-Axis 35 mm or 120 mm (depending on device type)
Integrated Variable Attenuation (Option) Filter wheel with 5 OD filters (OD 0–5)
Supply Data
ParameterSpecification
Power Supply 24 V DC ±5%, max. 1.8 A
Compressed Air for Cyclone 0.5–1 bar, clean, free of water and oil (ISO 8573-1:2010 class 6:4:4)
Cooling Water Pressure 500 W type only: 2 bar primary pressure with an unpressurized outflow, max. 4 bar
Min. Cooling Water Flow Rate 500 W type only: 1.5 l/min
Cooling Water Temperature Tin 500 W type only: dew point < Tin < 30 °C
Communication
ParameterSpecification
Interfaces Ethernet
Dimensions and Weight
ParameterSpecification
Dimensions (L × W × H) 427 mm (+12 mm traversing range + connector) × 202 mm × 181 mm (+35 mm / 120 mm traversing range + measuring objective projection)
Weight (Approx.) 15 kg
Features
Capabilities organized by functional domain.
Measurements

Measurements are carried out in conformity with ISO 11146.

  • Beam distribution in the individual measurement planes
  • Focal dimensions
  • Focal position in space
  • Rayleigh length
  • Divergence
  • Beam parameter product BPP
  • Diffraction index M²
How It Works

The MSM+ generates a magnified image of the incoming beam on a CCD sensor through its measuring objective, while additional attenuators and an optional filter wheel attenuate the beam gradually. The LaserDiagnosticsSoftware LDS calculates the beam diameter and position from the imaged power density distribution, and repeating the measurement along the working range determines the complete caustic. An integrated safety circuit monitors the equipment and shuts off the laser immediately in the event of a fault.

Operation

Two alternatives are available for operating and controlling the MSM+.

  • PC-based LaserDiagnosticsSoftware LDS for manual and semi-automatic measurement of the beam distribution in the focal area
  • Scripts for semi-automatic control of repetitive measurements in servicing, quality assurance and acceptance inspection
Measuring Objectives & Options

Three measuring objectives (MOB) are available — 3.3×, 5× and 15× — each coated for the NIR and optionally for the 2nd and 3rd harmonic.

  • Wavelength ranges 340, 360, 515, 545, 1,030 and 1,090 nm; others by adapting the lens coatings
  • Interchangeable fixed neutral density filters to attenuate peak powers of pulsed and USP lasers
  • Filter wheel with OD1–OD5 neutral density filters
  • Alternative beam radius definitions in LDS, including 2nd moments, 86% power inclusion, slit, knife-edge and Gauss fit
Request & Resources
View full Literature Library →
Website Contact Form