Short Beam Exposure

PocketMonitor

Overview

The PocketMonitor measures laser power according to the ballistic principle. In this method, an absorber is irradiated with laser radiation for a defined period (10 s/20 s). Based on the warming and the known weight of the absorber after a thermalization time, the irradiated power can be determined up to a resolution of 1 W (PMT 120icu) up to 0.01 W (PMT 002p).

25 W–12 kW (by model)
Power Range
±4%
Measuring Accuracy
10 s or 20 s
Irradiation Time
Specifications
PocketMonitor is offered in four models (PMT 05p / PMT 30p / PMT 70icu / PMT 120icu); values below are listed per model where they differ.
Measurement Parameters
ParameterSpecification
Models PMT 05p / PMT 30p / PMT 70icu / PMT 120icu
Power Range 05p: 25–500 W | 30p: 150–3,000 W | 70icu: 350–7,000 W | 120icu: 500–12,000 W
Wavelength Range 800–1,100 nm or 10.6 µm
Max. Beam Diameter on the Absorber 05p: 27.5 mm | 30p: 48 mm | 70icu/120icu: 36 mm
Max. Power Density on the Absorber 05p: 2.5 kW/cm² (<1 kW) | 30p: 2.5 kW/cm² (<1 kW), 1.5 kW/cm² (<3 kW) | 70icu/120icu: 5 kW/cm²
Irradiation Time 10 s at maximum permissible power, 20 s at 50% of maximum
Device Parameters
ParameterSpecification
Max. Angle of Incidence (Perpendicular to Inlet Aperture) ±5°
Max. Centered Tolerance ±2.0 mm
Measuring Accuracy ±4%
Reproducibility ±2%
Dimensions and Weight
ParameterSpecification
Absorber Diameter 05p: 45 mm | 30p: 79 mm | 70icu: 79 mm | 120icu: 99 mm
Absorber Height 05p: 15 mm | 30p: 20 mm | 70icu/120icu: 75 mm
Weight 05p: 0.56 kg | 30p: 0.67 kg | 70icu: 1.11 kg | 120icu: 1.55 kg
Request & Resources
View full Literature Library →
Website Contact Form