Short Beam Exposure

PowerMeasuringModule

Overview

The PowerMeasuringModule measures the laser power calorimetrically. The absorber is irradiated by the laser for a defined period. With the heat capacity of the test body, the temperature increase and the exposure time, the optical power is calculated in a few seconds. The exposure time measurement increases the reliability of the results. The measurement result is independent of the beam dimension and the beam position. The absorber cools via thermal conduction to the surrounding air. This restricts the measurement frequency of the system. The plant controller can query the absorber temperature at any time and determine the new irradiation period. A mechanical closing device at the inlet opening and protective glass protect the measuring device from dirt. The maximum power density on the absorber is 1.5 kW/cm².

400–12,000 W
Power Range
±3%
Accuracy
30 mm
Max. Beam Diameter
Specifications
Datasheet specifications for the PRIMES PowerMeasuringModule PMM.
Measurement Parameters
ParameterSpecification
Power Range 400–12,000 W (within the permitted maximum energy, E = P × t)
Wavelength Range 440–460 nm (BG), 510–540 nm (BG) and 800–1,100 nm (NIR)
Max. Beam Diameter 30 mm
Max. Power Density (Peak, ~25 mm Below the Protective Window) >10 mm beam: 4 kW/cm² | 10–3 mm: 5 kW/cm² | 3–1.5 mm: 10 kW/cm² | <1.5 mm: 12 kW/cm²
Irradiation Time (CW and Pulsed) 0.1–2.0 s depending on laser power (type AP3s: 0.1–1.0 s)
Pulse Length (Pulsed Operation) >50 µs: average and peak power, up to 10 kHz at 50% duty cycle | <50 µs: average power
Max. Laser Rise Time <1% of irradiation time
Energy per Measurement 400–3,500 J
Recommended Energy per Measurement 400–1,000 J
Total Duration Until Output <5 s (AP3s); <15 s for devices with the former thermalization standard
Nominal Measurement Frequency 500 J: 1 cycle/2 min | 1,000 J: 1 cycle/4 min
Device Parameters
ParameterSpecification
Max. Absorber Temperature 120 °C
Max. Angle of Incidence (Perpendicular to Inlet Aperture) ±10°
Accuracy (Angle of Incidence up to 10°) ±3%
Reproducibility ±1%
Supply Data
ParameterSpecification
Power Supply, DC In 24 V DC +25%/−20%; 250 mA
Power Supply, DC Out 24 V DC, max. 5 A
Communication
ParameterSpecification
Interfaces (Alternatively) PROFINET (copper/fiber optics), PROFIBUS, Parallel, DeviceNet, EtherNet/IP, EtherCAT
Dimensions and Weight
ParameterSpecification
Dimensions (L × W × H) 200 × 100 × 89 mm closed | 246 × 100 × 227 mm open
Weight (Approx.) 2.2 kg
Request & Resources
View full Literature Library →
Website Contact Form