Continuous Beam Exposure

PowerMonitor

Overview

The PRIMES PowerMonitor (PM) uses the same calorimetric principle as the CompactPowerMonitor. In contrast to the flat absorber in the CPM, the laser beam in the PowerMonitor is directed via a focusable mirror into a cylindrical absorber. A highly absorbent coating has been applied to the inside of the water-cooled absorber. The cylindrical absorber allows very high degrees of absorption with minimal back-reflection. This method is suited to very high powers. All parts that come into contact with the cooling water are manufactured from copper or brass. In this way stress corrosion is effectively prevented in the cooling cycle. A pneumatic closing device protects the PowerMonitor against dirt contamination.

PM 48: 0.3–8 kW / PM 100: 1–25 kW / PM HP: 3–150 kW
Power Range
±1%
Reproducibility
15 s up to 99% of final value
Time Constant
Specifications
PowerMonitor is offered in three models (PM 48 / PM 100 / PM HP); values below are listed per model where they differ.
Measurement Parameters
ParameterSpecification
Models PM 48 / PM 100 / PM HP
Power Range PM 48: 0.3–8 kW | PM 100: 1–25 kW | PM HP: 3–150 kW
Wavelength Range PM 48/100: 450 nm, 515–530 nm, 800–1100 nm, 2000 nm, 10600 nm | PM HP: 1000–1100 nm
Irradiation Time Continuous (all)
Max. Power Density PM 48 (450/515–532 nm): 10 kW/cm² | (800–1100 nm): 15 kW/cm² | PM 100 (all): 5 kW/cm² | PM HP (1000–1100 nm): 20 kW/cm²
Divergence (Full Angle) Min: −50 mrad | Max: +160 mrad (PM 48) / +180 mrad (PM 100 & HP)
Device Parameters
ParameterSpecification
Entrance Aperture PM 48: 48 mm | PM 100: 100 mm | PM HP: 90 mm
Accuracy ±2.0–3.5% depending on model and wavelength
Reproducibility ±1% (all)
Time Constant 15 s up to 99% of final value
Supply Data
ParameterSpecification
Power Supply 24 V ±5%, max. 0.5 A
Cooling Water Pressure (Min/Max) PM 48: 2/6 bar | PM 100: 1/4 bar | PM HP: 1/3 bar
Minimum Cooling Water Flow 0.5 l/min per kW (all)
Minimum Cooling Water Flow (Interlock) PM 48: 4 l/min | PM 100: 8 l/min | PM HP: 25 l/min
Maximum Cooling Water Flow PM 48: 12 l/min | PM 100: 30 l/min | PM HP: 150 l/min
Water Pressure Drop PM HP: 2.1 bar at 75 l/min
Cooling Water Temperature Tin Dew point < Tin < 30°C
Stability of Cooling Water Temperature <1 K/min or <0.08 K/5 s
Compressed Air Pressure (Min/Max) 2/4 bar
Compressed Air Purity Class ISO 8573-1:2010 [7:4:4]
Communication
ParameterSpecification
Interfaces RS485 / USB / Interlock / Analog out
Dimensions and Weight
ParameterSpecification
Dimensions (L × W × H) PM 48: 394 × 242 × 125 mm | PM 100: 580 × 330 × 215 mm | PM HP: 600 × 330 × 215 mm
Weight (Approx.) PM 48: 10 kg | PM 100: 44 kg | PM HP: 52 kg
Features
Capabilities organized by functional domain.
Design Requirements
  • Absorption of high-intensity beams
  • High degree of absorption
  • Long-term stability
  • Accuracy
  • Reproducibility
  • Short acquisition time
  • Reliable operation in tough operating environment
Request & Resources